Technical Support – Frequently Asked Questions
Is an upgrade available to make the PAD unit compatible with the AHA/ERC 2005 CPR Guidelines?
Yes. Please contact your AEDUniverse.com at 1-866-517-8243 or contact Heartsine Technical Support.
Status indicator is not flashing or flashing Red.
A problem has been detected. Check the expiry date of your Pad-Pak™. Change the Pad-Pak™ if the expiry date has been exceeded. If the status indicator is still not flashing or flashing red after you have changed your Pad-Pak™ contact AEDUniverse.com at 1-866-517-8243 or Heartsine technical support.
Does the PAD perform a weekly self test?
Yes this self test occurs on Sunday night at midnight GMT. If a fault is detected during self test the status indicator will flash red or will stop flashing. If this occurs please contact AEDUniverse.com at 1-866-517-8243 or HeartSine® technical support Heartsine Technical Support If the self test has passed the status indicator will continue to flash green every 5 seconds.
What test equipment can be used for testing the HeartSine® samaritan® PAD Defibrillator?
The only test equipment that can be used for testing the HeartSine® samaritan® PAD Defibrillator is the Fluke Biotek Defibrillator Tester QED 6 that has been calibrated after 1st June 2006. Please contact Heartsine technical support for further information.
Maintenance Checks Required - Watch a video
- Check the Status Indicator is flashing Green approximately every 5 seconds. If it is not flashing Green or flashing Red a fault has been detected. Change the Pad-Pak™. If the status indicator is still flashing Red or not flashing contact AEDUniverse.com at 1-866-517-8243.
- Check the expiration date of the Pad-Pak™ currently inserted into the PAD. If the Pad-Pak™ has exceeded the date stated on the lot number label, remove it from the PAD and replace with a new Pad-Pak™. Contact AEDUniverse.com at 1-866-517-8243 for replacements or click here to order replacements online.
- Check supplies, accessories and spares for damage or expiration. Replace any accessories found to be damaged or that have exceeded their expiration date.
- Check the exterior of the PAD for cracks or other signs of damage. Contact your AEDUniverse.com at 1-866-517-8243 if any damage is found.
- Check that trained responders are aware of the PADs location and that it is easily accessible for those responders at all times.
- Ensure all trained responders have up to date training for both CPR and AED use.
Warranty and Expiry
- Heartsine PAD AED – Extended 7 Year warranty for units shipped on or after 1 Oct 2006. Proof of date of Purchase required.
- PADs shipped prior to 1 October 2006 – 5 Year Warranty.
- Pad-Pak™ – 3.5 Year Shelf life from the date of manufacture.
Charge Time
Charging Time: 150J in less than 8 seconds, 200J in less than 12 seconds from a shock advised message.
Other Information
- PAD records 45 minutes of ECG and event/incident recording.
- Operating/Standby temperature: 0 to 50 C, 32 to 122 F
- Shipping/Transportation temperature: -10 to 50 C (14 to 122F) for up to two days. If device has been stored below 0C (32F) it should be returned to an ambient temperature of between 0 to 50C (32 to 122F) for at least 24 hours before use.
- Battery: 18V 0.8Amp Hours
(32 to 122F)
30 Shocks at 200J or 6 hours monitoring